ION-SELECTIVE ELECTRODES FOR ORGANIC CATIONS AND PHARMACEUTICALS (ISE, POLYMER MEMBRANE CALCIUM CHANNEL BLOCKERS).

Persistent Link:
http://hdl.handle.net/10150/187765
Title:
ION-SELECTIVE ELECTRODES FOR ORGANIC CATIONS AND PHARMACEUTICALS (ISE, POLYMER MEMBRANE CALCIUM CHANNEL BLOCKERS).
Author:
CUNNINGHAM, LAWRENCE JAMES.
Issue Date:
1984
Publisher:
The University of Arizona.
Rights:
Copyright © is held by the author. Digital access to this material is made possible by the University Libraries, University of Arizona. Further transmission, reproduction or presentation (such as public display or performance) of protected items is prohibited except with permission of the author.
Abstract:
Three areas related to Polymer Membrane Ion-Selective Electrodes were investigated. The first two concered (1) their application to analysis of organic cations and pharmaceuticals, and (2) the use of electrochemically polymerized films for potentiometric sensors. The third area focused on the development of a computer controlled data acquisition system for studying Ion Transport at the Interface of Two Immiscible Electrolyte Solutions (ITTIES). Coated wire Polymer Membrane Ion-Selective Electrodes (PMISEs) based on dinonylnapthalenesulfonic acid (DNNS) were prepared for several alkylammonium ions and pharmaceuticals. Selectivity coefficients and other operational characteristics were predictable from calculated distribution constants. Specifically, Log k ᵖᵒᵗ(i)(j) values increased linearly with carbon number in a homologous series of alkylammonium ions. Among cations of the same carbon number, interference increased tenfold with each additional substitution in going from quaternary, tertiary, secondary and primary alkylammonium ions. Addition of hydrophilic subtituent groups to cyclohexylammonium interferents further diminished electrode response. Electrodes for several important pharmaceuticals, namely cocaine, methamphetamine, methadone, protriptyline, cimetidine, lidoflazine, verapamil, acebutalol, diltiazem, and nicardipine, displayed nearly theoretical response down to micromolar concentrations. Their selectivity was consistent with calculated distribution constants. Selectivity was independent of sensitivity, suggesting that the magnitude of each was determined by the distribution of both the free cation and the association complex of this cation with DNNS. The polymerization of phenol derivatives onto copper rendered the resulting films responsive to counter-ions of the electroactive species present in the polymerization medium. Those incorporating Aliquat 336 (tricaprylmethylammonium chloride) showed response to various anions, while others prepared using a sulfonated phenol derivative (m-hydroxybenzenesulfonic acid) responded to tributylammonium. The films had short lifetimes due to poor adhesion to the copper substrate. Construction of a computer controlled data acquisition system provided a rapid method for obtaining electrocapillary data simultaneously with the current-scan polarogram. The system was evaluated by observation of ion-transfer of a variety of alkylammonium ions.
Type:
text; Dissertation-Reproduction (electronic)
Keywords:
Electrodes, Ion selective.; Electric apparatus and appliances.; Cations -- Analysis.; Drugs -- Analysis.; Thin films.
Degree Name:
Ph.D.
Degree Level:
doctoral
Degree Program:
Chemistry; Graduate College
Degree Grantor:
University of Arizona

Full metadata record

DC FieldValue Language
dc.language.isoenen_US
dc.titleION-SELECTIVE ELECTRODES FOR ORGANIC CATIONS AND PHARMACEUTICALS (ISE, POLYMER MEMBRANE CALCIUM CHANNEL BLOCKERS).en_US
dc.creatorCUNNINGHAM, LAWRENCE JAMES.en_US
dc.contributor.authorCUNNINGHAM, LAWRENCE JAMES.en_US
dc.date.issued1984en_US
dc.publisherThe University of Arizona.en_US
dc.rightsCopyright © is held by the author. Digital access to this material is made possible by the University Libraries, University of Arizona. Further transmission, reproduction or presentation (such as public display or performance) of protected items is prohibited except with permission of the author.en_US
dc.description.abstractThree areas related to Polymer Membrane Ion-Selective Electrodes were investigated. The first two concered (1) their application to analysis of organic cations and pharmaceuticals, and (2) the use of electrochemically polymerized films for potentiometric sensors. The third area focused on the development of a computer controlled data acquisition system for studying Ion Transport at the Interface of Two Immiscible Electrolyte Solutions (ITTIES). Coated wire Polymer Membrane Ion-Selective Electrodes (PMISEs) based on dinonylnapthalenesulfonic acid (DNNS) were prepared for several alkylammonium ions and pharmaceuticals. Selectivity coefficients and other operational characteristics were predictable from calculated distribution constants. Specifically, Log k ᵖᵒᵗ(i)(j) values increased linearly with carbon number in a homologous series of alkylammonium ions. Among cations of the same carbon number, interference increased tenfold with each additional substitution in going from quaternary, tertiary, secondary and primary alkylammonium ions. Addition of hydrophilic subtituent groups to cyclohexylammonium interferents further diminished electrode response. Electrodes for several important pharmaceuticals, namely cocaine, methamphetamine, methadone, protriptyline, cimetidine, lidoflazine, verapamil, acebutalol, diltiazem, and nicardipine, displayed nearly theoretical response down to micromolar concentrations. Their selectivity was consistent with calculated distribution constants. Selectivity was independent of sensitivity, suggesting that the magnitude of each was determined by the distribution of both the free cation and the association complex of this cation with DNNS. The polymerization of phenol derivatives onto copper rendered the resulting films responsive to counter-ions of the electroactive species present in the polymerization medium. Those incorporating Aliquat 336 (tricaprylmethylammonium chloride) showed response to various anions, while others prepared using a sulfonated phenol derivative (m-hydroxybenzenesulfonic acid) responded to tributylammonium. The films had short lifetimes due to poor adhesion to the copper substrate. Construction of a computer controlled data acquisition system provided a rapid method for obtaining electrocapillary data simultaneously with the current-scan polarogram. The system was evaluated by observation of ion-transfer of a variety of alkylammonium ions.en_US
dc.typetexten_US
dc.typeDissertation-Reproduction (electronic)en_US
dc.subjectElectrodes, Ion selective.en_US
dc.subjectElectric apparatus and appliances.en_US
dc.subjectCations -- Analysis.en_US
dc.subjectDrugs -- Analysis.en_US
dc.subjectThin films.en_US
thesis.degree.namePh.D.en_US
thesis.degree.leveldoctoralen_US
thesis.degree.disciplineChemistryen_US
thesis.degree.disciplineGraduate Collegeen_US
thesis.degree.grantorUniversity of Arizonaen_US
dc.identifier.proquest8424920en_US
dc.identifier.oclc691379732en_US
All Items in UA Campus Repository are protected by copyright, with all rights reserved, unless otherwise indicated.